Laboratory animals
-
The purpose of this study is to evaluate the accuracy of an oscillometric blood pressure monitor in anesthetized pigs. Invasive blood pressure (IBP) and noninvasive blood pressure (NIBP) measurements were taken using a DRE Waveline Pro multiparameter monitor at four different time points in 17 pigs undergoing injectable anesthesia. NIBP measurements were taken on both the thoracic and pelvic limbs. ⋯ Bias was significantly higher in the thoracic limb in comparison to the pelvic limb and was significantly higher at blood pressures above median. In general, NIBP measurements underestimated IBP measurements. In conclusion, the use of the DRE Waveline Pro to assess NIBP in anesthetized pigs may be useful in monitoring trends in mean and diastolic blood pressure and is most accurate when used on the pelvic limb.