Acute heart diseases are the most frequent causes for fatalities on merchant vessels. Presently there is no sufficient therapy available to treat ventricular fibrillation. The aim of this study was to test whether common automated external defibrillators [AED] may be appropriate for the use aboard merchant vessels. ⋯ All subjects handled the AED correctly. The AED received a total amount of points in the range between 2125 to 2241 (of 2400 possible). The subjects preferred AED with coloured as well as light marked buttons which gave a feedback (e.g. audible tones) when they were pressed. All AED were able to register an ECG in the vibrating ambient. Due to interface problems it was only possible to extract three ECG files, and only two files (data < 300 kB) could be sent as e-mail attachment via satellite to the German TMAS. In noisy areas the AED must guide the user, e.g. by screen massages and/or pictograms. Displays should provide additional data to help assess resuscitation effectiveness. A special procedure is necessary to ensure that ships and TMAS own the same software to read the transmitted ECG files, which are not allowed to exceed a size of 300 kB.