IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society
-
IEEE Trans Neural Syst Rehabil Eng · Mar 2007
Human nerve stimulation thresholds and selectivity using a multi-contact nerve cuff electrode.
Testing of the recruitment properties and selective activation capabilities of a multi-contact spiral nerve cuff electrode was performed intraoperatively in 21 human subjects. The study was conducted in two phases. An exploratory phase with ten subjects gave a preliminary overview of the data and data collection process and a systematic phase with eleven subjects provided detailed recruitment properties. ⋯ The selectivity, defined as the percent of total activation of the first muscle recruited before another muscle reached threshold, ranged from 27% to 97% with a mean of 55%. In each case, the muscle that was selectively activated was the first muscle to branch distal to the cuff location. This study serves as a preliminary evaluation of nerve cuff electrodes in humans prior to chronic implant in subjects with high tetraplegia.