A 128-channel 6 mW wireless neural recording IC with spike feature extraction and UWB transmitter.
This paper reports a 128-channel neural recording integrated circuit (IC) with on-the-fly spike feature extraction and wireless telemetry. The chip consists of eight 16-channel front-end recording blocks, spike detection and feature extraction digital signal processor (DSP), ultra wideband (UWB) transmitter, and on-chip bias generators. Each recording channel has amplifiers with programmable gain and bandwidth to accommodate different types of biological signals. ⋯ UWB telemetry is designed to wirelessly transfer raw data from 128 recording channels at a data rate of 90 Mbit/s. The chip is realized in 0.35 mum complementary metal-oxide-semiconductor (CMOS) process with an area of 8.8 x 7.2 mm(2) and consumes 6 mW by employing a sequential turn-on architecture that selectively powers off idle analog circuit blocks. The chip has been tested for electrical specifications and verified in an ex vivo biological environment.