Clinical biochemistry
-
Clinical biochemistry · Apr 1998
ReviewOn-line arterial blood gas analysis with optodes: current status.
To summarize the rationale for and the principles of blood gas and pH measurement with photochemical sensors (optodes) placed in the arterial line--either intravascularly (in vivo) or extravascularly (ex vivo). To review the specific problems that occur with in vivo measurement; the clinical data that have been obtained with continuous intravascular and on-demand extravascular systems; and, the role of this technology in the intensive care unit. ⋯ Before intravascular monitors can be used routinely for clinical care, reliability, consistency and accuracy will have to be demonstrated in large and widely divergent patient groups. Extravascular on-demand blood gas analysis is accurate, allows trend monitoring of blood gases and decreases the risk of infection, the therapeutic decision time and patient blood loss. As large patient studies are lacking the clinical role of on-line blood gas analysis cannot be clearly delineated.