Clinical biochemistry
-
Clinical biochemistry · Feb 2015
Using a simulation model to assess risk of false negative point-of-care urinary human chorionic gonadotropin device results due to high-dose hook interference.
It is unclear if the point-of-care (POC) Clinitest hCG device is subject to high-dose hook interference from physiological concentrations of intact human chorionic gonadotropin (hCG), β-core fragment of hCG (hCGβcf), and hCG free β-subunit (hCGβ) found in urine during pregnancy. We used a simulation model to address this question and related our findings to our institution's pregnant population in order to assess risk for potential false-negative hCG results. ⋯ The Clinitest-Clinitek Status hCG device is unlikely to exhibit false-negative urinary hCG results due to high-dose hook interference for women in early healthy pregnancy, although additional studies are necessary to determine potential risk in other patient populations. Visual interpretation of POC urinary hCG device results is an important failure mode to consider in risk analyses for erroneous urinary hCG device results.