Clinical biochemistry
-
Clinical biochemistry · Jan 2019
Multicenter StudyA novel Sigma metric encompasses global multi-site performance of 18 assays on the Abbott Alinity system.
The Abbott Alinity family of chemistry and immunoassay systems recently launched with early adopters contributing imprecision and bias data, which was consolidated to assess the performance of Alinity assays across multiple sites using the Sigma metric. Multi-site Sigma metrics were determined for 3 ion-selective electrodes, 12 photometric assays, and 3 immunoassays across 11 independent laboratory sites in 9 countries. ⋯ This multi-center study applies a novel application of Sigma metrics to the first Alinity users and reveals analytical performance of greater than 4 Sigma for vast majority of assays. Laboratories with limited resources can leverage larger data sets for Pooled Sigma metric analysis, providing a tool to assess the consistency of analytical performance from multiple sites.