• Respiratory care · Nov 2020

    Effect of Nitric Oxide Delivery Device on Tidal Volume Accuracy During Mechanical Ventilation at Small Tidal Volumes.

    • Courtney D Ranallo, Tracy L Thurman, Shirley J Holt, Summer G Frank-Pearce, Michael P Anderson, and Mark J Heulitt.
    • Department of Pediatrics/Critical Care Medicine, University of Oklahoma Health Sciences Center, Oklahoma City, Oklahoma. courtney-ranallo@ouhsc.edu.
    • Respir Care. 2020 Nov 1; 65 (11): 1641-1647.

    BackgroundInhaled nitric oxide (INO) is used in infants as a therapy for elevated pulmonary vascular resistance. When INO is delivered at low tidal volumes, displayed inspiratory and expiratory volumes vary widely. We hypothesize that volume is removed by the sampling line during the ventilation cycle, and this results in a net volume loss at low tidal volumes. This study aimed to measure the volumes delivered and to assess the accuracy of displayed ventilator values using a test lung.MethodsA test lung was connected to a ventilator and an INO delivery system. All tests were performed with stable mode settings across volumes of 18, 30, 42, and 60 mL. Flow measured with a pneumotachometer attached between the test lung and the circuit assessed the percent error between inspiratory and expiratory volumes measured by the pneumotachometer measured and displayed on the ventilator under various INO/sample line conditions to determine where and how much volume was being displaced.ResultsDisplayed and measured inspiratory volumes had small variations between the INO/sample line conditions and baseline. However, expiratory volumes, with the sample line connected, exhibited large percent error values that increased (-14, -20, -27, and -34) as tidal volume decreased (60, 42, 30, and 18 mL) and error was significantly larger compared to baseline in all tidal volumes (P < .01) with and without INO delivery.ConclusionsWe concluded that inspiratory volumes were not affected by INO delivery, but additional removal of volume in the expiratory phase of the breath cycle by the sampling line results in a large error in the displayed expiratory volume.Copyright © 2020 by Daedalus Enterprises.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.