• Neuromodulation · Apr 2003

    The performance and safety of an implantable spinal cord stimulation system in patients with chronic pain: a 5-year study.

    • Jean-Pierre Van Buyten.
    • Neuromodulation. 2003 Apr 1;6(2):79-87.

    AbstractReliability and ease of use of the Itrel 3 System (Medtronic Inc., Minneapolis, MN) were prospectively assessed over 5 years in patients with a range of pain syndromes (mainly low back and/or leg pain, or ischemic pain due to peripheral vascular disease). The longevity of the implantable pulse generator (IPG) battery, the frequency with which system settings were changed, and the ease of use of the EZ patient programmer were assessed. Data on adverse events, pain relief, and patient satisfaction with therapy were also collected. Following a screening procedure, 85 systems were implanted in 84 patients. Twenty-four patients were withdrawn prematurely and, in an additional 32 cases, end of battery life was reached before the end of the study. The survival curve for the IPG batteries showed that approximately 50% are expected to last up to the sixth year. No device failures or unanticipated device-related adverse events were reported. At least 90% of patients considered the EZ patient programmer easy to use. System settings were stable over time. The intensity and duration of pain were reduced significantly and patient satisfaction with therapy was high. We conclude that the Itrel 3 System performed well over 5 years and was easy to use. Its safety and effectiveness for the relief of chronic intractable pain of the trunk or limbs were also confirmed.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…