- Maria Daniela Falco, Domenico Genovesi, Luciana Caravatta, Clelia Di Carlo, Ekaterina Bliakharskaia, Marianna Appignani, Massimiliano Faustino, Nanda Furia, and Enrico Di Girolamo.
- Department of Radiation Oncology, "SS. Annunziata" Hospital, "G. D'Annunzio" University, Via De' Vestini, 66100, Chieti, Italy. mdanielafalco@hotmail.com.
- Strahlenther Onkol. 2021 Mar 1; 197 (3): 198-208.
Background And PurposeHigh-dose 6‑MV radiotherapy may cause cardiac implantable electronic devices (CIEDs) to malfunction. To assess CIED malfunctions resulting from direct exposure up to 10 Gy, 100 pacemakers (PMs) and 40 implantable cardioverter-defibrillators (ICDs) were evaluated.Materials And MethodsCIEDs underwent baseline interrogation. In ICDs, antitachycardia therapies were disabled via the programmer while the detection windows were left enabled. A computed tomography (CT) scan was performed to build the corresponding treatment plan. CIEDs were "blinded" and randomized to receive single doses of either 2, 5, or, 10 Gy via a 6-MV linear accelerator (linac) in a water phantom. Twenty-two wireless telemetry-enabled CIEDs underwent a real-time session, and their function was recorded by the video camera in the bunker. The CIEDs were interrogated after exposure and once monthly for 6 months.ResultsDuring exposure, regardless of dose, 90.9% of the CIEDs recorded electromagnetic interference, with 6 ICDs (27.3%) reporting pacing inhibition and inappropriate arrhythmia detections. After exposure, a backup reset was observed in 1 PM (0.7% overall, 1% among PMs), while 7 PMs (5% overall, 7% among PMs) reported battery issues (overall immediate malfunction rate was 5.7%). During follow-up, 4 PMs (2.9% overall; 4% among PMs) and 1 ICD (0.7% overall; 2.5% among ICDs) reported abnormal battery depletion, and 1 PM (0.7% overall; 1% among PMs) reported a backup reset (overall late malfunction rate was 4.3%).ConclusionApart from transient electromagnetic interference, last-generation CIEDs withstood direct 6‑MV exposure up to 10 Gy. Permanent battery or software errors occurred immediately or later only in less recent CIEDs.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.