• Heart Rhythm · May 2012

    Unpredictable battery depletion of St Jude Atlas II and Atlas+ II HF implantable cardioverter-defibrillators.

    • Cevher Ozcan, Jeffrey N Rottman, E Kevin Heist, Mary L Guy, Patrick T Ellinor, Jagmeet Singh, David J Milan, Stephan B Danik, Conor D Barrett, Moussa Mansour, Jeremy N Ruskin, and Theofanie Mela.
    • Cardiac Arrhythmia Service, Massachusetts General Hospital, Boston, MA, USA.
    • Heart Rhythm. 2012 May 1; 9 (5): 717-20.

    BackgroundPredictable progression to battery depletion is necessary for device management in patients with pacemakers or implantable cardioverter-defibrillators, particularly in patients who either are pacemaker dependent or have required implantable cardioverter-defibrillator therapies.ObjectiveTo determine the incidence and characteristics of unexpected battery depletion in patients implanted with a cardiac resynchronization therapy - defibrillator (CRT-D) device.MethodsAll patients with a St Jude Atlas+ HF or Atlas II HF CRT-D device implanted between 2004 and 2007 at the Massachusetts General Hospital and the Nashville VA Medical Center (Vanderbilt University) were studied. All patients with early generator depletion (transition of generator voltage above specified elective replacement indicator [ERI] to end of life [EOL] in less than 90 days) were evaluated further.ResultsEight cases (mean age 69.6 ± 9 years) with abrupt battery depletion were identified among 191 patients (4.2%) implanted with a St Jude Atlas CRT-D device. The longevity of 8 premature depletion devices was 46.4 ± 10 months (median 45 months). The battery voltage in these 8 devices decreased from a mean of 2.48 ± 0.03 V (above ERI) to 2.3 ± 0.08 V (below ERI) over 33.3 ± 23 days (range 1-59 days; median 38.5 days). One device reached EOL status within 1 day of having battery voltage above ERI and another device within 12 days.ConclusionThe incidence of abrupt battery depletion was 4.2% in patients implanted with a St Jude Atlas CRT-D device. No common mechanism has been identified for this failure. Close monitoring of battery voltage and timely generator replacement are required in patients with these devices.Copyright © 2012 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.