    Lack of relationship between the true airway pressure and the pressure displayed with an infant ventilator.

    • A Sola, D Fariña, S Rodriguez, and I Kurlat.
    • Division of Neonatology, Hospital de Pediatría J.P. Garrahan, Buenos Aires, Argentina.
    • Crit. Care Med. 1992 Jun 1; 20 (6): 778-81.

    ObjectiveTo establish if the pressure indicated in the manometer of an infant ventilator (IV 100B, Sechrist, Anaheim, CA) reflects the true pressure delivered to the proximal airway during mechanical ventilation in the neonatal ICU.DesignWith approval of our Institutional Research Board, data were collected prospectively. Peak inspiratory pressure and end-expiratory pressure were measured at the "Y" piece of the breathing tubing. Pressure readings from the conventional ventilator's manometer were compared with simultaneously obtained measurements using an electronic monitor.SettingThis study was conducted in a 45-bed neonatal ICU, admitting 700 to 750 newborns per year.PatientsTwelve neonates who required mechanical ventilation were included in the study.InterventionsSpecific interventions were not made by study design. Measurements routinely obtained were compared.Measurements And Main ResultsTwo hundred seventy-five simultaneous measurements of peak inspiratory pressure and positive end-expiratory pressure were compared. Peak inspiratory pressure values were higher with the electronic monitor in 273 (99%) of 275 measurements and the mean of the differences between the electronic monitor and ventilator's manometer was statistically significant (p less than .001). For positive end-expiratory pressure measurements, values indicated by the electronic monitor were lower in 152 (55%) of 275 determinations, equal in 65 (23%), and higher in 58 (21%) determinations. Percent variations between methods ranged from 0% to 140% for peak inspiratory pressures and from 0% to 500% for positive end-expiratory pressure.ConclusionsThese data demonstrate that it is impossible to know the true pressure delivered to the proximal airway of a neonate during mechanical ventilation by observing the ventilator pressure manometer. The manometer readings consistently underestimate the true peak inspiratory pressure values and are very unpredictable regarding positive end-expiratory pressure values. These findings support the use of other methods to monitor the proximal airway pressure besides the ventilator's manometer in the neonatal ICU. Furthermore, mean airway pressure should not be calculated from the pressure readings obtained from the tested ventilator's manometer.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.