• Artificial organs · Apr 2018

    Comparative Study

    In Vitro Comparison of Two Neonatal ECMO Circuits Using a Roller or Centrifugal Pump With Three Different In-Line Hemoconcentrators for Maintaining Hemodynamic Energy Delivery to the Patient.

    • Madison Force, Morgan Moroi, Shigang Wang, David A Palanzo, Allen R Kunselman, and Akif Ündar.
    • Department of Pediatrics, Penn State Health Pediatric Cardiovascular Research Center, Penn State College of Medicine, Penn State Health Children's Hospital, Hershey, PA, USA.
    • Artif Organs. 2018 Apr 1; 42 (4): 354-364.

    AbstractThe objective of this study was to compare three different hemoconcentrators (Hemocor HPH 400, Mini, and Junior) with two different neonatal ECMO circuits using a roller or a centrifugal pump at different pseudo-patient pressures and flow rates in terms of hemodynamic properties. This evidence-based research is necessary to optimize the ECMO circuitry for neonates. The circuits used a 300-mL soft-shell reservoir as a pseudo-patient approximating the blood volume of a 3 kg neonate, two blood pumps, and a Quadrox-iD Pediatric oxygenator with three different in-line hemoconcentrators (Hemocor HPH 400, Mini, and Junior). One circuit used a Maquet H20 roller pump and another circuit used a Maquet RotaFlow centrifugal pump. The circuit was primed with lactated Ringer's solution followed by heparinized packed red blood cells with a hematocrit of 40%. The pseudo-patient's pressure was manually maintained at 40, 60, or 80 mm Hg and the flow rate was maintained at 200, 400, or 600 mL/min with a circuit temperature of 36°C. Pressure and flow data was recorded using a custom-made data acquisition device. Mean pressures, diverted blood flow, pressure drops, and total hemodynamic energy (THE) were calculated for each experimental condition. The roller pump and centrifugal pump performed similarly for all hemodynamic properties with all experimental conditions. The Hemocor HPH Junior hemoconcentrator added the highest resistance to the circuit. The Hemocor HPH Junior provided the highest circuit pressures, lowest diverted blood flow, highest pressure drop across the circuit, and highest THE generated by the pump. The Hemocor HPH 400 added the least resistance to the circuit, providing the lowest circuit pressures, more diverted flow, lowest pressure drop, and the lowest THE generated by the pump. However, the THE delivered to the patient was the same for the three hemoconcentrators. While the three hemoconcentrators performed differently in terms of hemodynamic properties throughout the circuit, the THE transmitted to the patient was similar for all three hemoconcentrators due to the consistent pseudo-patient's pressure that was manually maintained for each trial. While the THE delivered to the patient indicates similar perfusion for these patients with any of the three hemoconcentrators, the differences in added resistance to the circuit may impact the decision of which hemoconcentrator is used. There was no clinically significant difference between the two circuits with the roller versus centrifugal pump in terms of hemodynamic properties in this study. Further in vivo research is warranted to confirm our findings.© 2018 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…