J Int Soc Respir Prot · Jan 2017
Development of a Manikin-Based Performance Evaluation Method for Loose-Fitting Powered Air-Purifying Respirators.
- Mike Bergman, Rohan Basu, Zhipeng Lei, George Niezgoda, and Ziqing Zhuang.
- National Institute for Occupational Safety and Health, National Personal Protective Technology Laboratory, Pittsburgh, Pennsylvania 15236.
- J Int Soc Respir Prot. 2017 Jan 1; 34 (1): 40-57.
ObjectiveLoose-fitting powered air-purifying respirators (PAPRs) are increasingly being used in healthcare. NIOSH has previously used advanced manikin headforms to develop methods to evaluate filtering facepiece respirator fit; research has now begun to develop methods to evaluate PAPR performance using headforms. This preliminary study investigated the performance of PAPRs at different work rates to support development of a manikin-based test method.MethodsManikin penetration factors (mPF) of three models of loose-fitting PAPRs were measured at four different work rates (REST: 11 Lpm, LOW: 25 Lpm, MODERATE: 48 Lpm, and HIGH: 88 Lpm) using a medium-sized NIOSH static advanced headform mounted onto a torso. In-mask differential pressure was monitored throughout each test. Two condensation particle counters were used to measure the sodium chloride aerosol concentrations in the test chamber and also inside the PAPR facepiece over a 2-minute sample period. Two test system configurations were evaluated for returning air to the headform in the exhalation cycle (filtered and unfiltered). Geometric mean (GM) and 5th percentile mPFs for each model/work rate combination were computed. Analysis of variance tests were used to assess the variables affecting mPF.ResultsPAPR model, work rate, and test configuration significantly affected PAPR performance. PAPR airflow rates for the three models were approximately 185, 210, and 235 Lpm. All models achieved GM mPFs and 5th percentile mPFs greater than their designated Occupational Safety and Health Administration assigned protection factors despite negative minimum pressures observed for some work rate/model combinations.ConclusionsPAPR model, work rate, and test configuration affect PAPR performance. Advanced headforms have potential for assessing PAPR performance once test methods can be matured. A manikin-based inward leakage test method for PAPRs can be further developed using the knowledge gained from this study. Future studies should vary PAPR airflow rate to better understand the effects on performance. Additional future research is needed to evaluate the correlation of PAPR performance using advanced headforms to the performance measured with human subjects.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.