Differentiating False Loss of Resistance from True Loss of Resistance While Performing the Epidural Block with the CompuFlo® Epidural Instrument.
- Pasquale Vaira, Michela Camorcia, Tiziana Palladino, Matteo Velardo, and Giorgio Capogna.
- Department of Anesthesiology, Casa Sollievo della Sofferenza Hospital, San Giovanni Rotondo, (FG), Italy.
- Anesthesiol Res Pract. 2019 Jan 1; 2019: 5185901.
BackgroundThe occurrence of false losses of resistance may be one of the reasons for inadequate or failed epidural block. A CompuFlo® epidural instrument has been introduced to measure the pressure of human tissues in real time at the orifice of a needle and has been used as a tool to identify the epidural space. The aim of this study was to investigate the sensitivity and the specificity of the ability of CompuFlo® to differentiate the false loss of resistance from the true loss of resistance encountered during the epidural space identification procedure.MethodWe performed epidural block with the CompuFlo® epidural instrument in 120 healthy women who requested labor epidural analgesia. The epidural needle was considered to have reached the epidural space when an increase in pressure (accompanied by an increase in the pitch of the audible tone) was followed by a sudden and sustained drop in pressure for more than 5 seconds accompanied by a sudden decrease in the pitch of the audible tone, resulting in the formation of a low and stable pressure plateau. We evaluate the sensitivity, specificity, and positive and negative predictive values of the ability of CompuFlo® recordings to correctly identify the true LOR from the false LOR.ResultsThe drop in pressure associated with the epidural space identification was significantly greater than that recorded after the false loss of resistance (73% vs 33%) (P=0.000001). The sensitivity was 0.83, and the AUC was 0.82.DiscussionWe have confirmed the ability of CompuFlo® to differentiate the false loss of resistance from the true loss of resistance and established its specificity and sensitivity.ConclusionAn easier identification of dubious losses of resistance during the epidural procedure is essential to reduce the number of epidural attempts and/or needle reinsertions with the potential of a reduced risk of accidental dural puncture especially in difficult cases or when the procedure is performed by trainees.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.