Extrusion of expandable stacked interbody device for lumbar fusion: case report of a complication.
- Paul D Kim, Eli M Baron, and Michel Levesque.
- Spine Institute of San Diego, San Diego, CA, USA.
- Spine. 2012 Aug 15;37(18):E1155-8.
Study DesignA case report.ObjectiveWe report on the device failure of a polyetherether-ketone expandable cage device with posterior migration of one of its components.Summary Of Background DataPosterior migration of transforaminal lumbar interbody fusion cage devices has been reported with static devices, and ours is the first report of an expandable TLIF device failure and posterior migration of one of its components.MethodsThe patient is a 30-year-old man who had previously failed 3 lumbar surgical procedures and presented for L5-S1 lumbar fusion with pedicle screws and transforaminal interbody fusion.ResultsPostoperative imaging demonstrated posterior migration of one of the failed expandable interbody components with eventual revision surgery and placement of static transforaminal lumbar interbody fusion cages.ConclusionThis is the first case report to describe such complication, and caution must be warranted when using these devices.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.