J Vasc Interv Radiol · Oct 2006
Comparative StudyControl and performance characteristics of eight different suction biopsy devices.
- Randy R Sibbitt, Wilmer L Sibbitt, Sharon E Nunez, Lawrence G Kettwich, Sharon C Kettwich, and Arthur D Bankhurst.
- Department of Radiology, St. Peter's Hospital, Helena, Montana, USA. wsibbitt@salud.unm.edu
- J Vasc Interv Radiol. 2006 Oct 1; 17 (10): 1657-69.
PurposeTo determine the control and performance characteristics of eight different suction biopsy devices.Materials And MethodsPhysician control of the syringe and needle was measured precisely with the validated linear displacement method during the aspiration phase and during five biopsy passes. The visual analog scale was used to measure operator difficulty in the following domains: (i) attachment to the needle, (ii) generation of vacuum, (iii) detection of loss of vacuum, (iv) release of the vacuum, and (v) clearing of the sample from the needle.ResultsPerformance in various phases of the biopsy procedure varied widely among the biopsy devices tested. Unintended forward penetration (ie, loss of control in the forward direction) was significant with the reverse aspiration syringe (31.5 +/- 1.7 mm), three-ringed control syringe (25.4 +/- 4.1 mm), BioSuc-C7 syringe (28.3 +/- 1.9 mm), conventional syringe with a plunger lock (6.1 +/- 1.5 mm), syringe pistol (9.2 +/- 2.4 mm), and conventional syringe (3.8 +/- 2.9 mm) but was significantly less for the reciprocating procedure device (RPD; 0.7 +/- 0.7 mm; P = .001) and RPD syringe holder (0.6 +/- 0.6 mm; P = .001). The mean performance rankings were the best for the RPD (3.42 +/- 2.57) and RPD syringe holder (4.29 +/- 2.50) and worst for the conventional syringe (6.14 +/- 2.67; P = .001) and conventional syringe with a plunger lock (6.86 +/- 3.80; P = .001).ConclusionsEach of the suction biopsy devices has unique advantages and disadvantages. Suction biopsy devices with the least favorable overall performance were the conventional syringe and the conventional syringe with a plunger lock. The highest overall performance was seen with the RPD and RPD syringe holder.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.