• Clin Physiol Funct Imaging · Nov 2005

    Clinical Trial

    Recording of dynamic arterial compliance changes during hand elevation.

    • Kersti Jagomägi, Rein Raamat, Jaak Talts, Ulvi Ragun, Esko Länsimies, and Jukka Jurvelin.
    • Institute of Physiology, University of Tartu, Tartu, Estonia. kersti.jagomagi@ut.ee
    • Clin Physiol Funct Imaging. 2005 Nov 1; 25 (6): 350-6.

    AbstractFinger arterial compliance has been studied on the beat-to-beat basis by using the digital arterial pressure and volume waveforms and performing measurements at zero transmural pressure during arm elevation in 11 volunteers. Continuous non-invasive finger blood pressure was measured by applying the Finapres monitor and the finger volume pulses were recorded by the UT9201 physiograph by using the photoplethysmographic principle of registration. Estimation of the beat-to-beat finger arterial compliance is based on (i) the recorded volume and pressure wave amplitudes (Vpulse and Ppulse) and (ii) on the calculation of the slope of the pressure-volume relationship from the first derivatives dV/dt and dP/dt of the recorded volume and pressure pulses near the point of the maximum slope. The results of the study demonstrate that the applied two methods similarly (correlation coefficient r = 0.97) describe the changes of the beat-to-beat compliance during hand elevation test. At the same time the second estimate was 18% higher than the first one (P = 0.003).

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        