• Childs Nerv Syst · Mar 2017

    Ventriculoperitoneal shunt malfunction caused by fractures and disconnections over 10 years of follow-up.

    • Fatih Serhat Erol, Sait Ozturk, Bekir Akgun, and Metin Kaplan.
    • School of Medicine, Department of Neurosurgery, Firat University, Elazig, Turkey.
    • Childs Nerv Syst. 2017 Mar 1; 33 (3): 475-481.

    PurposeWe have discussed the diagnosis and treatment approaches in patients with discontinued (disconnected or fractured) ventriculoperitoneal (VP) shunts that caused mechanical dysfunction.MethodsBetween 2006 and 2016, VP shunt surgery was performed on 1357 pediatric patients in our clinic. In follow-up examinations, we retrospectively reviewed patients who underwent revision surgery. Except for diagnosis of discontinued VP shunt, by excluding revision-surgery patients, only those patients who underwent surgical treatment owing to discontinued (fracture or disconnection) catheter were included in the study. Age at first surgery, sex, reason for shunt discontinuity, anatomical region of pathology, time to discontinuity diagnosis after first surgery, and presence or absence of symptoms were evaluated.ResultsOne thousand three hundred fifty-seven VP shunt surgeries were performed in total, with 305 (22.4%) patients requiring revision surgery. Of these 305 patients, after accounting for other complications like obstruction, infection, overdrainage, and so on, 25 (8.1%) patients (14 male, 11 female) required re-surgery due to discontinuity. The mean age of these patients was 5.4 ± 2.1 months during the first VP shunt surgery, and the mean age during revision surgery was 71.7 months. The mean duration until discontinuity was diagnosed was 66.3 ± 24.1 months (76.1 months for catheter fractures and 45.6 months for disconnections (p 0.021)).ConclusionDisconnection and fracture are two significant mechanical VP shunt dysfunctions and must be adequately researched and understood even during routine follow-ups. A disconnected or fractured shunt may be working and it is not safe to state that the shunt is no longer needed.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…