• Ann Pharmacother · Jul 1999

    Comparative Study Clinical Trial

    Accuracy, clinical correlation, and patient acceptance of two handheld prothrombin time monitoring devices in the ambulatory setting.

    • D C Chapman, M A Stephens, G L Hamann, L E Bailey, and C S Dorko.
    • The Regional Medical Center at Memphis, TN 38103, USA.
    • Ann Pharmacother. 1999 Jul 1; 33 (7-8): 775-80.

    ObjectiveTo evaluate the accuracy, clinical correlation, ease of use, and patient acceptance of the Coaguchek and the ProTime Microcoagulation System as compared with standard laboratory methods for prothrombin time determination.MethodsA total of 30 prothrombin times, expressed as international normalized ratios (INRs), were determined by each handheld device for comparison with standard laboratory testing. Accuracy was evaluated by calculating the absolute difference for each pair of INR values. Clinical correlation was defined as an INR obtained by the handheld monitor that would have resulted in the same therapeutic decision as the INR obtained by the standard laboratory method. Subjects were surveyed to determine which method of INR determination they preferred and their reasons for that preference.ResultsAccuracy was superior with the Coaguchek monitor. The absolute difference (mean +/- SD) in the laboratory and Coaguchek INRs was 0.28+/-0.23 (p = 0.96). The absolute difference (mean +/- SD) in the laboratory and the ProTime Microcoagulation System INRs was 0.56+/-0.34 (p < 0.001). For clinical correlation, two out of 24 (8.3%) INRs with the Coaguchek were sufficiently different from the laboratory INR to have resulted in a different therapeutic decision, compared with 12 out of 24 (50%) with the ProTime Microcoagulation System (p < 0.005). Of subjects surveyed, 77.8% preferred the finger stick method.ConclusionsThe Coaguchek was superior to the ProTime Microcoagulation System in accuracy, clinical correlation, and ease of use. The study also showed that patients preferred capillary blood sampling by finger puncture over venipuncture for INR monitoring.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…