• Lasers Surg Med · Mar 2012

    MR-based thermometry of laser induced thermotherapy: temperature accuracy and temporal resolution in vitro at 0.2 and 1.5 T magnetic field strengths.

    • Thomas J Vogl, Frank Huebner, Nagy N N Naguib, Ralf W Bauer, Martin G Mack, Nour-Eldin A Nour-Eldin, and Dirk Meister.
    • Institute for Diagnostic and Interventional Radiology, Johann Wolfgang Goethe University Hospital, Frankfurt University, Frankfurt, Germany. t.vogl@em.uni-frankfurt.de
    • Lasers Surg Med. 2012 Mar 1; 44 (3): 257-65.

    PurposeTo evaluate MR-thermometry using fast MR sequences for laser induced interstitial thermotherapy (LITT) at 0.2 and 1.5 T systems.Methods & MaterialsIn-vitro experiments were performed using Agarose gel mixture and lobes of porcine liver. MR-thermometry was performed by means of longitudinal relaxation time (T1) and proton resonance frequency shift (PRF) methods under acquisition of amplitude and phase shift images. Four different sequences were used for T1 thermometry: A gradient-echo (GRE), a True Fast Imaging with Steady Precession (TRUFI), a Saturation Recovery Turbo-FLASH (SRTF), and an Inversion Recovery Turbo-FLASH (IRTF) sequence (FLASH-Fast Low Angle Shot). PRF was measured with four sequences: Two fast-spoiled GRE sequences (one as WIP sequence), a Turbo-FLASH (TFL) sequence (WIP sequence), and a multiecho-TrueFISP sequence. Temperature was controlled and verified using a fiber-optic Luxtron device. The temperature was correlated with the MR measurement.ResultsAll sequences showed a good linear correlation R(2) = 0.97-0.99 between the measured temperature and the MR-thermometry measurements. The only exception was the TRUFI sequence in the Agarose phantom that showed a non-linear calibration curve R(2) = 0.39-0.67. At 1.5 T, the Agarose experiments revealed similar temperature accuracies of 4-6°C for all sequences excluding TRUFI. During experiments with the liver, the PRF sequences showed better performance than the T1, with accuracies of 5-12°C, contrary to the T1 sequences at 14-18°C. The accuracy of the Siemens PRF-FLASH sequence was 5.1°C. At 0.2 T, the Agarose experiments provided the highest accuracy of 3.3°C for PRF measurement. At the liver experiments the T1 sequences SRTF and FLASH revealed the best accuracies at 6.4 and 7.0°C.ConclusionThe accuracy and speed of MR temperature measurements are sufficient for controlling the temperature-based tumor destruction. For 0.2 T systems SRTF and FLASH sequences are recommended. For 1.5 T systems SRTF and FLASH are the most accurate.Copyright © 2012 Wiley Periodicals, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…