• NeuroImage · May 2020

    Comparative Study

    Test-retest reliability and sample size estimates after MRI scanner relocation.

    • Tracy R Melzer, Ross J Keenan, Gareth J Leeper, Stephen Kingston-Smith, Simon A Felton, Sarah K Green, Kerry J Henderson, Nickolas J Palmer, Reza Shoorangiz, Mustafa M Almuqbel, and Daniel J Myall.
    • Department of Medicine, University of Otago, Christchurch, New Zealand; New Zealand Brain Research Institute, Christchurch, New Zealand; Brain Research New Zealand - Rangahau Roro Aotearoa Centre of Research Excellence, New Zealand. Electronic address: tracy.melzer@otago.ac.nz.
    • Neuroimage. 2020 May 1; 211: 116608.

    ObjectiveMany factors can contribute to the reliability and robustness of MRI-derived metrics. In this study, we assessed the reliability and reproducibility of three MRI modalities after an MRI scanner was relocated to a new hospital facility.MethodsTwenty healthy volunteers (12 females, mean age (standard deviation) ​= ​41 (11) years, age range [25-66]) completed three MRI sessions. The first session (S1) was one week prior to the 3T GE HDxt scanner relocation. The second (S2) occurred nine weeks after S1 and at the new location; a third session (S3) was acquired 4 weeks after S2. At each session, we acquired structural T1-weighted, pseudo-continuous arterial spin labelled, and diffusion tensor imaging sequences. We used longitudinal processing streams to create 12 summary MRI metrics, including total gray matter (GM), cortical GM, subcortical GM, white matter (WM), and lateral ventricle volume; mean cortical thickness; total surface area; average gray matter perfusion, and average diffusion tensor metrics along principal white matter pathways. We compared mean MRI values and variance at the old scanner location to multiple sessions at the new location using Bayesian multi-level regression models. K-fold cross validation allowed identification of important predictors. Whole-brain analyses were used to investigate any regional differences. Furthermore, we calculated within-subject coefficient of variation (wsCV), intraclass correlation coefficient (ICC), and dice similarity index (SI) of cortical segmentations across scanner relocation and within-site. Additionally, we estimated sample sizes required to robustly detect a 4% difference between two groups across MRI metrics.ResultsAll global MRI metrics exhibited little mean difference and small variability (bar cortical gray matter perfusion) both across scanner relocation and within-site repeat. T1- and DTI-derived tissue metrics showed ​< ​|0.3|% mean difference and <1.2% variance across scanner location and <|0.4|% mean difference and <0.8% variance within the new location, with between-site intraclass correlation coefficient (ICC) ​> ​0.80 and within-subject coefficient of variation (wsCV) ​< ​1.4%. Mean cortical gray matter perfusion had the highest between-session variability (6.7% [0.3, 16.7], estimate [95% uncertainty interval]), and hence the smallest ICC (0.71 [0.44,0.92]) and largest wsCV (13.4% [5.4, 18.1]). No global metric exhibited evidence of a meaningful mean difference between scanner locations. However, surface area showed evidence of a mean difference within-site repeat (between S2 and S3). Whole-brain analyses revealed no significant areas of difference between scanner relocation or within-site. For all metrics, we found no support for a systematic difference in variance across relocation sites compared to within-site test-retest reliability. Necessary sample sizes to detect a 4% difference between two independent groups varied from a maximum of n ​= ​362 per group (cortical gray matter perfusion), to total gray matter volume (n ​= ​114), average fractional anisotropy (n ​= ​23), total gray matter volume normalized by intracranial volume (n ​= ​19), and axial diffusivity (n ​= ​3 per group).ConclusionCortical gray matter perfusion was the most variable metric investigated (necessitating large sample sizes to identify group differences), with other metrics showing substantially less variability. Scanner relocation appeared to have a negligible effect on variability of the global MRI metrics tested. This manuscript reports within-site test-retest variability to act as a tool for calculating sample size in future investigations. Our results suggest that when all other parameters are held constant (e.g., sequence parameters and MRI processing), the effect of scanner relocation is indistinguishable from within-site variability, but may need to be considered depending on the question being investigated.Copyright © 2020 The Authors. Published by Elsevier Inc. All rights reserved.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.