• J Am Heart Assoc · Feb 2019

    Novel Left Heart Catheterization Ramp Protocol to Guide Hemodynamic Optimization in Patients Supported With Left Ventricular Assist Device Therapy.

    • Andrew N Rosenbaum, Robert P Frantz, Sudhir S Kushwaha, John M Stulak, Simon Maltais, and Atta Behfar.
    • 1 Department of Cardiovascular Diseases Mayo Clinic Rochester MN.
    • J Am Heart Assoc. 2019 Feb 19; 8 (4): e010232.

    AbstractBackground Left ventricular (LV) hemodynamic assessment has been sparsely performed in patients supported on continuous-flow LV assist devices (cf LVADs ). Insight into dynamic changes of left heart parameters during ramp studies may improve LV assist device optimization and evaluate pathology. Methods and Results To complement right heart catheterization, a novel technique for left heart catheterization in patients with a cf LVAD was developed. Patients implanted with cf LVAD s underwent hemodynamic ramp left heart catheterization and right heart catheterization with transthoracic echocardiography. Continuous aortic and LV pressures were measured along with right atrial, pulmonary artery, and pulmonary capillary wedge pressures. A novel index, the transaortic gradient ( TAG ) was established. Thirty eight patients with cf LVADs were evaluated at a median of 446 days (interquartile range, 183-742) after device implant. During left heart catheterization performed for hemodynamic optimization, drop-in LV end-diastolic pressure and pulmonary capillary wedge pressure were associated with a rise in TAG . A range was identified for TAG (20-40 mm Hg) as providing the most optimal level of hemodynamic offloading. Pathologic states deviated from normal responses to ramp. LV assist device thrombosis was associated with an inability to increase in TAG during speed ramp. Significant aortic insufficiency was associated with a marked increase in LV end-diastolic pressure despite a concomitant decrease in pulmonary capillary wedge pressure with increasing LV assist device speeds. Conclusions Inclusion of left heart catheterization to a typical right heart catheterization LV assist device ramp protocol imparted unique insights to optimize cf LVAD speeds in different clinical scenarios. A novel index, the TAG was defined and provided additional resolution to optimized offloading.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…