AJR Am J Roentgenol · Nov 2015
Comparative StudyComparison of Standardized Uptake Values in Normal Structures Between PET/CT and PET/MRI in a Tertiary Pediatric Hospital: A Prospective Study.
- Karen Lyons, Victor Seghers, James I L Sorensen, Wei Zhang, Michael J Paldino, Rajesh Krishnamurthy, and Eric M Rohren.
- 1 Department of Radiology, Texas Children's Hospital, 6701 Fannin St, Ste 470, Houston, TX 77030.
- AJR Am J Roentgenol. 2015 Nov 1; 205 (5): 1094-101.
ObjectiveThe purpose of this study was to compare standardized uptake values (SUVs) of normal tissues using MR attenuation-corrected versus CT attenuation-corrected (18)F-FDG PET in a pediatric population.Subjects And MethodsThirty-five patients (21 boys; mean age, 13.3 years) referred for 47 PET/CT scans were recruited to undergo PET/MRI. MR attenuation correction was performed using an automated three-segment model. ROIs were drawn over nine normal structures to estimate SUV(min), SUV(mean), and SUV(max). Pearson rank correlation coefficients were calculated to compare SUVs obtained from MR and CT attenuation correction. In nine patients who underwent multiple PET/MRI studies, coefficients of variance and intraclass correlation coefficients were calculated to evaluate intrapatient SUV(max) variation.ResultsMean (± SD) time to imaging after FDG injection was 108 ± 17 minutes for PET/CT and 61 ± 6 minutes for PET/MRI. PET/MRI SUVs in all tissues were lower than those for PET/CT (mean difference, -28.9% ± 31.1%; p < 0.05). Very high or high correlation between PET/MRI and PET/CT SUV(max) was found in brain (r = 0.72), myocardium (r = 0.95), and bone marrow (r = 0.85) (p < 0.001). Moderate correlation was found in liver (r = 0.54), fat (r = 0.41), mean blood pool (r = 0.40), and psoas muscle (r = 0.38) (p < 0.01). Weak correlation was found in lung (r = 0.12) and iliacus muscle (r = 0.12). Compared with PET/CT, PET/MRI systematically undermeasured SUV. In nine patients who underwent multiple PET/MRI examinations, moderate or strong agreement was found in the SUV(max) of six of nine tissues, similar to the corresponding PET/CT examinations.ConclusionOur study showed overall high correlation for SUV measurements obtained from MR attenuation correction compared with CT attenuation correction, although PET/MRI underestimated SUV compared with PET/CT. SUVs measured from PET/MRI indicated good intrapatient reliability.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.