• J Clin Anesth · Nov 1997

    Case Reports

    False low pulse oximetry reading associated with the concomitant use of a peripheral nerve stimulator and an evoked-potential stimulator.

    • I Keidan, A Sidi, and D Gravenstein.
    • Department of Anesthesiology, University of Florida College of Medicine, Gainesville 32610-0254, USA.
    • J Clin Anesth. 1997 Nov 1; 9 (7): 591-6.

    AbstractOne of the sources of error in pulse oximetry readings is associated with an abnormal signal-to-noise ratio. The pulse oximeter distinguishes the light absorbance of arterial blood from that of other absorbers by differentiating between a constant component and a pulsating component. The pulsating component is almost exclusively the result of arteriolar bed pulsations. Because pulse oximetry is based on the assumption that arterial blood is the only pulsatile absorber, any other fluctuating phenomenon could constitute a source of error. We report a case in which a low pulse oximetry reading was associated with concomitant use of a pulse oximeter and a peripheral nerve stimulator on the same arm. Further tests conducted using a nerve stimulator and a sensory evoked potential stimulator with different amplitudes and frequencies confirmed the association and delineated the relationship between frequency and amplitude of stimulation and the degree of artificial desaturation. A theoretical explanation for this phenomenon is presented.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…