• Comput Methods Programs Biomed · May 2003

    Comparative Study Clinical Trial

    A program based on a 'selective' least-squares method for respiratory mechanics monitoring in ventilated patients.

    • André Eberhard, Pierre-Yves Carry, Jean-Pierre Perdrix, Jean-Marc Fargnoli, Loïc Biot, and Pierre F Baconnier.
    • Laboratoire de Modélisation et Calcul, Institut IMAG, 51 rue des Mathématiques, BP 53, 38041 Grenoble Cedex 9, France.
    • Comput Methods Programs Biomed. 2003 May 1; 71 (1): 39-61.

    AbstractThis paper proposes a program for continuous estimation of respiratory mechanics parameters in ventilated patients. This program can be used with any ventilator providing airway pressure and flow signals without additional equipment. Overall breathing resistance, dynamic elastance (E) and positive end expiratory pressure (P(0)) are periodically estimated by multiple linear regression on selected parts of breathing cycles. Experimental validation together with justification of the selection procedure are based on signals obtained while ventilating a lung mechanical analogue with various intensive care ventilators. Clinical validity has been tested on 12 ventilated patients. The quality of estimation has been assessed by mean square difference between measured and reconstituted pressure (MSE), coefficient of determination (R(2)) and the condition number (a confidence index), and by comparison of E and P(0) with corresponding static values. The high R(2) and the low MSE obtained on most clinical cycles indicate that selected parts of cycles obey closely the model underlying parameter estimation. Agreement between static and dynamic parameters demonstrates the clinical validity of our program.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…