-
Case Reports
Occlusion of the pilot tube in a Laser-Shield II endotracheal tube caused by methylene blue crystals: a case report.
- T Hashimoto and V E Armstead.
- Department of Anesthesiology, Jefferson Medical College, Thomas Jefferson University, Philadelphia, PA 19107-5587, USA.
- J Clin Anesth. 1998 Sep 1; 10 (6): 522-3.
AbstractThe Laser-Shield II (Xomed-Treace, Jacksonville, FL) endotracheal tube is equipped with methylene blue crystals to aid in the prompt detection of tube cuff rupture. As the tube cuff is being inflated with saline, the crystals are supposed to dissolve fully in the saline, forming a solution that is readily visible in the event of cuff rupture. We describe a case of occlusion of the pilot balloon tube caused by undissolved methylene blue crystals, a situation that made it impossible to deflate the Laser-Shield II tube cuff. Our experience may aid in future modification of endotracheal tube design.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.