• Plos One · Jan 2020

    Comparative Study

    In vitro comparison of performance including imposed work of breathing of CPAP systems used in low-resource settings.

    • Megan Heenan, Jose D Rojas, Z Maria Oden, and Rebecca Richards-Kortum.
    • Rice 360°: Institute for Global Health, Rice University, Houston, Texas, United States of America.
    • Plos One. 2020 Jan 1; 15 (12): e0242590.

    AbstractRespiratory distress due to preterm birth is a significant cause of death in low-resource settings. The introduction of continuous positive airway pressure (CPAP) systems to treat respiratory distress significantly reduced mortality in high-resource settings, but CPAP was only recently introduced in low-resource settings due to cost and infrastructure limitations. We evaluated pressure stability and imposed work of breathing (iWOB) of five CPAP systems used in low resource settings: the Fisher and Paykel bubble CPAP, the Diamedica baby CPAP, the Medijet nCPAP generator, and the first (2015) and second (2017) generation commercially available Pumani CPAPs. Pressure changes due to fresh gas flow were evaluated for each system by examining the relationship between flow and pressure at the patient interface for four pressures generated at the bottle (0, 3, 5, and 7 cm H2O); for the Medijet nCPAP generator, no bottle was used. The slope of the resulting relationship was used to calculate system resistance. Poiseuille's law of resistance was used to investigate significant contributors to resistance. Resistance ranged from 0.05 to 1.40 [Formula: see text]; three CPAP devices had resistances < 0.4 [Formula: see text]: the Fisher and Paykel system, the Diamedica system, and the second generation Pumani bubble CPAP. The other two systems, the Medijet nCPAP generator and the first generation Pumani bCPAP, had resistances >1.0 [Formula: see text]. Imposed WOB was measured using an ASL5000 test lung to simulate the breath cycle for an infant (5.5 kg), a term neonate (4.0 kg), and a preterm neonate (2.5 kg). Imposed WOB ranged from 1.4 to 39.5 mJ/breath across all systems and simulated infant sizes. Changes in pressure generated by fresh gas flow, resistance, and iWOB differ between the five systems evaluated under ideal laboratory conditions. The available literature does not indicate that these differences affect clinical outcomes.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…