Oncology nursing forum · Oct 1996
Comparative Study Clinical Trial Controlled Clinical TrialThe effects of heparin flush on patency of the Groshong catheter: a pilot study.
- D J Mayo, M K Horne, B L Summers, D C Pearson, and C B Helsabeck.
- Warren Grant Magnuson Clinical Center of the National Institutes of Health in Bethesda, MD, USA.
- Oncol Nurs Forum. 1996 Oct 1; 23 (9): 1401-5.
Purpose/ObjectivesTo determine whether the addition or a heparinized saline flush would decrease clot formation and persistent withdrawal occlusion (PWO) in Groshong (Bard Access Systems, Salt Lake City, UT) catheters.DesignA prospective, nonrandomized study using a historical control group of patients with Groshong catheters that had been flushed weekly with 5 ml normal saline compared to data from patients with Groshong catheters flushed weekly with 2.5 ml heparinized saline (100 U/ml). A retrospective chart review was performed to determine the incidence of PWO. In both groups, the presence of liquid blood and adherent or nonadherent clot in explanted catheters was recorded.SettingOncology inpatient and outpatient units of a cancer research center located in a mid-Atlantic city in the United States.SampleControl GroupTwenty-eight 9.5 Fr. double-lumen Groshong catheters maintained with a saline flush. Experimental group: Twenty-three double-lumen Groshong catheters maintained with a heparin flush. At the time of this report, 12 of these 23 catheters had been explanted and 11 remained in place.MethodsThe frequency of PWO was measured in a retrospective chart review and determined by the number of urokinase instillations required for each catheter. All 28 catheters in the saline flush group and 8 catheters in the heparin flush group were examined immediately after removal for intraluminal liquid blood and adherent or nonadherent clot.Main Research VariablesUrokinase usage, intra-luminal blood or clot, and PWO.FindingsPWO occurred less frequently in the heparin flush group (p = 0.006) than in the saline fush group. All 28 of the saline flush catheters developed an adherent clot in one or both lumens, whereas no adherent clots were found in the heparin flush catheters (p < 0.0001).ConclusionsThe addition of a heparinized saline weekly flush to maintain Groshong catheters decreased the presence of intraluminal adherent clots and improved the catheter function.Implications For Nursing PracticeThe safe delivery of medication and the ability to obtain blood specimens are vital for patients who depend on functioning venous access catheters. Flushing Groshong catheters with heparinized saline decreases the likelihood of intraluminal clot formation and catheter malfunction.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.