Comput Methods Programs Biomed · Jul 2020
Glucose outcomes of a learning-type artificial pancreas with an unannounced meal in type 1 diabetes.
- Lulu Song, Changqing Liu, Wenying Yang, Jinping Zhang, Xiaomu Kong, Bo Zhang, Xiaoping Chen, Na Wang, Dong Shen, Zhaoqing Li, Xian Jin, Ying Shuai, and Youqing Wang.
- Department of Endocrinology, China-Japan Friendship Hospital, Beijing, China.
- Comput Methods Programs Biomed. 2020 Jul 1; 191: 105416.
Background And ObjectivesGlycemic control with unannounced meals is the major challenge for artificial pancreas. In this study, we described the performance and safety of learning-type model predictive control (L-MPC) for artificial pancreas challenged by an unannounced meal in type 1 diabetes (T1D).MethodsThis closed-loop (CL) system was tested in 29 T1D patients at one site in a 4 h inpatient open-label study. Participants used an L-MPC CL system for 6 days after 2-day system identification using open-loop (OL) insulin system. During the CL period, the L-MPC system was started from 8:00 am to noon each day. At 9:00 am, each participant consumed 50 g of carbohydrates with no prandial insulin bolus. At 9:30 am on CL-Day 4 or CL-Day 6, participants rode bicycles for 20 minutes or drank 50 ml of beer, in a random order.ResultsAs the primary outcome, TIR on CL-Day 3 was 65.2±23.3%, which was 9.8 points higher (95% CI 1.8 to 17.8; P = 0.019) than that on CL-Day 1. The time of glucose >10 mmol/L was decreased by 11.0% (95% CI -18.7 to 3.3; P = 0.007), and mean glucose level was decreased by 1.1 mmol/L (95% CI -1.1 to 0.5; P = 0.000). The total daily insulin dosage showed no significant difference (-0.1U, 95% CI -1.34 to 1.32; P = 0.982). Compared with OL-Day1 with a postprandial bolus, the TIR was increased by 13.7 points (95% CI 1.4 to 26.0; P = 0.030), the time of glucose >10 mmol/L and the mean glucose level were also decreased. Compared with the exercise day (CL-Day E, 62.0 ± 23.3%; P = 0.347) or alcohol day (CL-Day A, 64.0 ± 23.6%; P = 0.756), there was no statistically significant difference in terms of TIR, time of glucose >10 mmol/L and mean glucose level. No severe hypoglycemic events occurred and hypoglycemic episodes were not increased by using closed-loop insulin system.ConclusionThe L-MPC CL insulin system achieved good glycemic control challenged by an unannounced meal.Copyright © 2020 Elsevier B.V. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.