• Clinical biochemistry · Jul 2019

    Analytical characterization of the Siemens Dimension EXL high-sensitivity cardiac troponin I assay.

    • Peter A Kavsak, Jackie MacCuish, Jill Boreyko, Chantele Roy, Shana Lamers, and Lorna Clark.
    • Department of Pathology and Molecular Medicine, McMaster University, Hamilton, ON, Canada; Core Laboratory, Hamilton Health Sciences, Hamilton, ON, Canada. Electronic address: kavsakp@mcmaster.ca.
    • Clin. Biochem. 2019 Jul 1; 69: 52-56.

    BackgroundSiemens Healthcare Diagnostics has four commercially available assays on different analytical platforms using different methodologies to generate signal. We assessed the analytical performance of the Dimension EXL hs-cTnI assay (LOCI method) across different matrices and compared it to two different acridinium ester-based hs-cTnI assays (ADVIA Centaur and Abbott ARCHITECT).MethodsThe analytical sensitivity and precision below the 99th-percentile was determined for the Dimension EXL hs-cTnI assay. Method comparisons were performed between the Dimension EXL contemporary cTnI and the hs-cTnI assays, between different matrices for the EXL hs-cTnI assay (serum, lithium heparin and EDTA plasma), and between different hs-cTnI assays (EXL versus ADVIA Centaur or Abbott ARCHITECT) using non-parametric analyses.ResultsThe limit of blank and detection were 0.9 ng/L and 1.7 ng/L, respectively, with imprecision of 5.8% at 8.6 ng/L and 3.2% at 47.5 ng/L. Comparison between the EXL contemporary cTnI and hs-cTnI assay (range: 2.6-4214 ng/L) yielded proportional lower concentrations for the hs-cTnI assay (slope = 0.86; 95%CI: 0.81 to 0.96, n = 40); however, there was no difference in concentrations below 100 ng/L between the assays (median difference = -2.7 ng/L; 95%CI: -9.8 to 9.3). Passing-Bablok regression analysis with EDTA plasma yielded proportionally higher concentrations with the EXL hs-cTnI versus Abbott hs-cTnI (slope = 1.45; 95%CI: 1.02-1.86, n = 40) with proportionally lower concentrations with EDTA versus lithium heparin plasma with the EXL hs-cTnI assay alone (slope = 0.93; 95%CI: 0.90 to 0.99, n = 40). Comparison with Abbott hs-cTnI concentrations below 100 ng/L in the three matrices, indicated that the EXL hs-cTnI assay yielded higher concentrations (median difference range: 3.4-9.4 ng/L), with differences also evident when comparing the EXL hs-cTnI assay to the ADVIA Centaur hs-cTnI assay.ConclusionThe Siemens EXL hs-cTnI assay meets the analytical criteria for a high-sensitivity assay, with assay specific cutoffs important to maximize clinical performance.Copyright © 2019 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.