Randomized Controlled Trial Multicenter StudyEfficacy and safety of a parylene-coated occluder for atrial septal defect: a prospective, multi-center, randomized controlled clinical trial.
- Kai Yang, Jian-Hua Lyu, Hai-Bo Hu, Wen-Hui Wu, Ge-Jun Zhang, Jing-Lin Jin, Yu-Hao Liu, Yuan Feng, Li-Ming Wang, Shi-Hua Zhao, Shi-Liang Jiang, Wei Li, and Xiang-Bin Pan.
- Center of Structural Heart Disease, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences, Peking Union Medical College, Beijing 100037, China.
- Chin. Med. J. 2021 Nov 10; 134 (22): 2685-2691.
BackgroundNitinol-containing devices are widely used in clinical practice. However, there are concerns about nickel release after nitinol-containing device implantation. This study aimed to compare the efficacy and safety of a parylene-coated occluder vs. a traditional nitinol-containing device for atrial septal defect (ASD).MethodsOne-hundred-and-eight patients with ASD were prospectively enrolled and randomly assigned to either the trial group to receive a parylene-coated occluder (n = 54) or the control group to receive a traditional occluder (n = 54). The plugging success rate at 6 months after device implantation and the pre- and post-implantation serum nickel levels were compared between the two groups. A non-inferiority design was used to prove that the therapeutic effect of the parylene-coated device was non-inferior to that of the traditional device. The Cochran-Mantel-Haenszel chi-squared test with adjustment for central effects was used for the comparison between groups.ResultsAt 6 months after implantation, successful ASD closure was achieved in 52 of 53 patients (98.11%) in both the trial and control groups (95% confidence interval (CI): [-4.90, 5.16]) based on per-protocol set analysis. The absolute value of the lower limit of the 95% CI was 4.90%, which was less than the specified non-inferiority margin of 8%. No deaths or severe complications occurred during 6 months of follow-up. The serum nickel levels were significantly increased at 2 weeks and reached the maximum value at 1 month after implantation in the control group (P < 0.05 vs. baseline). In the trial group, there was no significant difference in the serum nickel level before vs. after device implantation (P > 0.05).ConclusionsThe efficacy of a parylene-coated ASD occluder is non-inferior to that of a traditional uncoated ASD occluder. The parylene-coated occluder prevents nickel release after device implantation and may be an alternative for ASD, especially in patients with a nickel allergy.Copyright © 2021 The Chinese Medical Association, produced by Wolters Kluwer, Inc. under the CC-BY-NC-ND license.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.