Multicenter Study Observational Study
Comparison of neuromuscular block measured by compressomyography at the upper arm and electromyography at the adductor pollicis muscle in obese and non-obese patients: An observational study.
- G H M Honing, B Torensma, C H Martini, K B Hoek, M van Velzen, E Olofsen, M Niesters, A Dahan, and M Boon.
- Department of Anesthesiology LUMC, Albinusdreef 2, 2333, ZA, Leiden, the Netherlands. Electronic address: g.h.m.honing@lumc.nl.
- J Clin Anesth. 2022 Jun 1; 78: 110673.
Study ObjectiveQuantitative neuromuscular monitoring is traditionally evaluated at the adductor pollicis muscle. By contrast, the TOF-Cuff compressomyograph evaluates neuromuscular block (NMB) at the upper arm. However, compressomyography has not been fully validated against other monitoring entities. This study evaluates the agreement between NMB measured by compressomyography at the upper arm and electromyography at the adductor pollicis muscle during various levels of neuromuscular block in patients with and without obesity.InterventionsNMB was measured at the upper arm by compressomyography (TOF-Cuff) and by electromyography (GE-NMT) at the adductor pollicis.DesignProspective, multicenter, observational study.SettingSecondary and tertiary care hospitals' operating theatres.Patients200 non-obese and 50 obese patients.MeasurementsDuring onset and offset of deep (post-tetanic-count 1-15 twitches), moderate (Train-of-Four-count 1-3 twitches) and shallow (Train-of-Four-ratio 0.01-1.0) depths of NMB were measured in obese and non-obese patients. The bias and limits of agreement of both devices were calculated using a Bland-Altman analysis for repeated measurements. Data obtained during spontaneous recovery (i.e. without the use of reversal agents) were used in the primary analyses.Main ResultsData from enrolled patients yielded 942 paired post-tetanic-counts, 1175 paired train-of-four-counts and 1574 paired train-of-four ratios during spontaneous recovery. In non-obese patients, mean bias (95% CI) between the two devices was 3.405 (2.294 to 4.517) during deep NMB; -0.023 (-0.205 to 0.160) during moderate NMB and 0.312 (0.287 to 0.338) during shallow NMB. In obese patients, bias was -0.170 (-2.872 to 2.531); 0.178 (-0.202 to 0.558); 0.384 (0.299 to 0.469) for deep, moderate and shallow NMB respectively.ConclusionsThere is variable disagreement between the level of NMB measured at the upper arm by compressomyography and at the adductor pollicis muscle measured by electromyography, throughout the various stages of NMB in obese and non-obese patients. Recovery of NMB on compressomyography preceded recovery on electromyography, which may have consequences for reversal and extubation decisions in clinical practice.Copyright © 2022 The Authors. Published by Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.