• Chinese medical journal · Jul 2018

    Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements.

    • Hai-Jian Zhao, Meng-Lei Ge, Yin Yan, Tian-Jiao Zhang, Jie Zeng, Wei-Yan Zhou, Yu-Fei Wang, Qing-Hui Meng, and Chuan-Bao Zhang.
    • National Center for Clinical Laboratories, Beijing Hospital, National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Beijing 100730, China.
    • Chin. Med. J. 2018 Jul 5; 131 (13): 1584-1590.

    BackgroundMeasuring total serum calcium is important for the diagnosis of diseases. Currently, results from commercial kits for calcium measurement are variable. Generally, the performance of serum calcium measurements is monitored by external quality assessment (EQA) or proficiency testing schemes. However, the commutability of the EQA samples and calibrators is often unknown, which limits the effectiveness of EQA schemes. The aim of this study was to evaluate the bias of serum calcium measurements and the commutability of processed materials.MethodsInductively coupled plasma mass spectrometry was applied as a comparative method, and 14 routine methods were chosen as test methods. Forty-eight serum samples from individual patients and 25 processed materials were quantified. A scatter plot was generated from patient samples, and 95% prediction intervals were calculated to evaluate the commutability of the processed materials and measurement bias at three concentration levels was used to determine the accuracy of routine assays.ResultsAll assays showed high precision (total coefficient of variation [CV] <2.26%) and correlation coefficients (r > 0.99). For all assays, the mean bias for the 48 patient samples ranged from -0.13 mmol/L to 0.00 mmol/L (-5.61-0.01%), and the ranges for the three concentrations were -0.10-0.04 mmol/L (-5.71-2.35%), -0.14--0.01 mmol/L (-5.80--0.30%), and -0.19-0.04 mmol/L (-6.24-1.22%). The EQA samples, calibrators, and animal sera exhibited matrix effects in some assays; human serum pools were commutable in all assays; certificate reference materials were commutable in most assays, and only GBW09152 exhibited a matrix effect in one assay; and aqueous reference materials exhibited matrix effects in most assays.ConclusionsBiases for most assays were within the acceptable range, although the accuracy of some assays needs improvement. Human serum pools prepared from patient samples were commutable, and the other tested materials exhibited a matrix effect.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…