• A&A practice · Jul 2022

    Case Reports

    Fire Hazard in an Anesthesia Machine: A Case Report.

    • Suman Sokhal, Navdeep Sokhal, Pradeep Sharma, and Rajendra Singh Chouhan.
    • From the Department of Neuroanaesthesiology and Critical Care, All India Institute of Medical Sciences, New Delhi, India.
    • A A Pract. 2022 Jul 1; 16 (7): e01603.

    AbstractAn operating room (OR) fire is a rare event but may cause critical harm to patients and health care personnel. We present 2 fire incidents in the rear cabinet of an anesthesia machine, not previously reported in the literature. An anesthesia machine in standby mode is considered safer than in use, but in the first case, the fire occurred while the machine was on standby mode. In another case, the fire erupted while surgery was in progress. Here, we discuss the management of an anesthesia machine fire and emphasize the need for improved fire safety in advanced anesthesia machines.Copyright © 2022 International Anesthesia Research Society.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..