- Suman Sokhal, Navdeep Sokhal, Pradeep Sharma, and Rajendra Singh Chouhan.
- From the Department of Neuroanaesthesiology and Critical Care, All India Institute of Medical Sciences, New Delhi, India.
- A A Pract. 2022 Jul 1; 16 (7): e01603.
AbstractAn operating room (OR) fire is a rare event but may cause critical harm to patients and health care personnel. We present 2 fire incidents in the rear cabinet of an anesthesia machine, not previously reported in the literature. An anesthesia machine in standby mode is considered safer than in use, but in the first case, the fire occurred while the machine was on standby mode. In another case, the fire erupted while surgery was in progress. Here, we discuss the management of an anesthesia machine fire and emphasize the need for improved fire safety in advanced anesthesia machines.Copyright © 2022 International Anesthesia Research Society.