• Ann. Intern. Med. · Dec 2022

    Randomized Controlled Trial Multicenter Study

    Perioperative Safety and Early Patient and Device Outcomes Among Subcutaneous Versus Transvenous Implantable Cardioverter Defibrillator Implantations : A Randomized, Multicenter Trial.

    • Jeff S Healey, Andrew D Krahn, Jamil Bashir, Guy Amit, François Philippon, William F McIntyre, Bernice Tsang, Jacqueline Joza, Derek V Exner, David H Birnie, Mouhannad Sadek, Darryl P Leong, Markus Sikkel, Victoria Korley, John L Sapp, Jean-Francois Roux, Shun Fu Lee, Gloria Wong, Angie Djuric, Danna Spears, Sandra Carroll, Eugene Crystal, Tom Hruczkowski, Stuart J Connolly, Blandine Mondesert, and ATLAS Investigators.
    • Population Health Research Institute, Hamilton, Ontario, Canada (J.S.H., W.F.M., D.P.L., S.F.L., G.W., A.D., S.J.C.).
    • Ann. Intern. Med. 2022 Dec 1; 175 (12): 165816651658-1665.

    BackgroundImplantable cardioverter defibrillators (ICDs) improve survival in patients at risk for cardiac arrest, but are associated with intravascular lead-related complications. The subcutaneous ICD (S-ICD), with no intravascular components, was developed to minimize lead-related complications.ObjectiveTo assess key ICD performance measures related to delivery of ICD therapy, including inappropriate ICD shocks (delivered in absence of life-threatening arrhythmia) and failed ICD shocks (which did not terminate ventricular arrhythmia).DesignRandomized, multicenter trial. (ClinicalTrials.gov: NCT02881255).SettingThe ATLAS trial.Patients544 eligible patients (141 female) with a primary or secondary prevention indication for an ICD who were younger than age 60 years, had a cardiogenetic phenotype, or had prespecified risk factors for lead complications were electrocardiographically screened and 503 randomly assigned to S-ICD (251 patients) or transvenous ICD (TV-ICD) (252 patients). Mean follow-up was 2.5 years (SD, 1.1). Mean age was 49.0 years (SD, 11.5).MeasurementsThe primary outcome was perioperative major lead-related complications.ResultsThere was a statistically significant reduction in perioperative, lead-related complications, which occurred in 1 patient (0.4%) with an S-ICD and in 12 patients (4.8%) with TV-ICD (-4.4%; 95% CI, -6.9 to -1.9; P = 0.001). There was a trend for more inappropriate shocks with the S-ICD (hazard ratio [HR], 2.37; 95% CI, 0.98 to 5.77), but no increase in failed appropriate ICD shocks (HR, 0.61 (0.15 to 2.57). Patients in the S-ICD group had more ICD site pain, measured on a 10-point numeric rating scale, on the day of implant (4.2 ± 2.8 vs. 2.9 ± 2.2; P < 0.001) and 1 month later (1.3 ± 1.8 vs. 0.9 ± 1.5; P = 0.035).LimitationAt present, the ATLAS trial is underpowered to detect differences in clinical shock outcomes; however, extended follow-up is ongoing.ConclusionThe S-ICD reduces perioperative, lead-related complications without significantly compromising the effectiveness of ICD shocks, but with more early postoperative pain and a trend for more inappropriate shocks.Primary Funding SourceBoston Scientific.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…