    Comparison of Four Mechanical Insufflation-Exsufflation Devices: Effect of Simulated Airway Collapse on Cough Peak Flow.

    • Nicolas Terzi, Isabelle Vaugier, Claude Guérin, Hélène Prigent, Ghilas Boussaid, Karl Leroux, Mathieu Delorme, Frédéric Lofaso, and Bruno Louis.
    • Centre Hospitalier Universitaire Grenoble Alpes, Médecine Intensive Réanimation, INSERM, Université Grenoble-Alpes, Grenoble, France. nicolas.terzi@chu-rennes.fr.
    • Respir Care. 2023 Apr 1; 68 (4): 462469462-469.

    BackgroundMechanical insufflation-exsufflation (MI-E) devices are used to improve airway clearance in individuals with acute respiratory failure. Some MI-E devices measure cough peak flow (CPF) during MI-E to optimize pressure adjustments. The aim was to compare CPF and effective cough volume (ECV: volume expired/coughed > 3 L/s) measurements between 4 MI-E devices under simulated conditions of stable versus collapsed airway.MethodsFour MI-E devices were tested on the bench. Each device was connected via a standard circuit to a collapsible tube placed in an airtight chamber that was attached to a lung model with adjustable compliance and resistance. Pressure was measured upstream and downstream the collapsing tube; air flow was measured between the chamber and the lung model. Each device was tested in 2 conditions: collapse condition (0 cm H2O) and no-collapse condition (-70 cm H2O). For each condition, 6 combinations of inspiratory/expiratory pressures were applied. CPF was measured at the "mouth level" by the device built-in flow meter and at the "tracheal level" by a dedicated pneumotachograph. Comparisons were performed with non-parametric tests.ResultsCPF values measured at the tracheal level and ECV values differed between devices for each inspiratory/expiratory pressure in the collapse and no-collapse conditions (P < .001). CPF values were significantly lower at the tracheal level in the collapse as compared with the no-collapse condition (P < .001 for each device), whereas they were higher at the mouth level (P < .05) for 3 of the 4 devices.ConclusionsCPF values differed significantly across MI-E devices, highlighting limitation(s) of using only CPF values to determine cough effectiveness. In simulated of airway collapse, CPF increased at the mouth, whereas it decreased at the tracheal level.Copyright © 2023 by Daedalus Enterprises.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…