- Nicolas Terzi, Isabelle Vaugier, Claude Guérin, Hélène Prigent, Ghilas Boussaid, Karl Leroux, Mathieu Delorme, Frédéric Lofaso, and Bruno Louis.
- Centre Hospitalier Universitaire Grenoble Alpes, Médecine Intensive Réanimation, INSERM, Université Grenoble-Alpes, Grenoble, France. nicolas.terzi@chu-rennes.fr.
- Respir Care. 2023 Apr 1; 68 (4): 462469462-469.
BackgroundMechanical insufflation-exsufflation (MI-E) devices are used to improve airway clearance in individuals with acute respiratory failure. Some MI-E devices measure cough peak flow (CPF) during MI-E to optimize pressure adjustments. The aim was to compare CPF and effective cough volume (ECV: volume expired/coughed > 3 L/s) measurements between 4 MI-E devices under simulated conditions of stable versus collapsed airway.MethodsFour MI-E devices were tested on the bench. Each device was connected via a standard circuit to a collapsible tube placed in an airtight chamber that was attached to a lung model with adjustable compliance and resistance. Pressure was measured upstream and downstream the collapsing tube; air flow was measured between the chamber and the lung model. Each device was tested in 2 conditions: collapse condition (0 cm H2O) and no-collapse condition (-70 cm H2O). For each condition, 6 combinations of inspiratory/expiratory pressures were applied. CPF was measured at the "mouth level" by the device built-in flow meter and at the "tracheal level" by a dedicated pneumotachograph. Comparisons were performed with non-parametric tests.ResultsCPF values measured at the tracheal level and ECV values differed between devices for each inspiratory/expiratory pressure in the collapse and no-collapse conditions (P < .001). CPF values were significantly lower at the tracheal level in the collapse as compared with the no-collapse condition (P < .001 for each device), whereas they were higher at the mouth level (P < .05) for 3 of the 4 devices.ConclusionsCPF values differed significantly across MI-E devices, highlighting limitation(s) of using only CPF values to determine cough effectiveness. In simulated of airway collapse, CPF increased at the mouth, whereas it decreased at the tracheal level.Copyright © 2023 by Daedalus Enterprises.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.