- Michael McPeck, Jane Moon, Jeyanthan Jayakumaran, and Gerald C Smaldone.
- Pulmonary, Critical Care and Sleep Medicine Division, Department of Medicine, Stony Brook University Medical Center, Stony Brook, New York.
- Respir Care. 2023 Sep 1; 68 (9): 121312201213-1220.
BackgroundTo understand the fate of aerosols delivered by high-flow nasal cannula using continuous nebulization, an open-source anatomical model was developed and validated with a modified real-time gamma ratemeter technique. Mass balance defined circuit losses. Responsiveness to infusion rate and device technology were tested.MethodsA nasal airway cast derived from a computed tomography scan was converted to a 3-dimensional-printed head and face structure connected to a piston ventilator (breathing frequency 30 breaths/min, tidal volume 750 mL, duty cycle 0.50). For mass balance experiments, saline mixed with Technetium-99m was infused for 1 h. Aerosol delivery was measured using a gamma ratemeter oriented to an inhaled mass filter at the hypopharynx of the model. Background and dead-space effects were minimized. All components were imaged by scintigraphy. Continuous nebulization was tested at infusion rates of 10-40 mL/h with gas flow of 60 L/min using a breath-enhanced jet nebulizer (BEJN), and a vibrating mesh nebulizer. Drug delivery rates were defined by the slope of ratemeter counts/min (CPM/min) versus time (min).ResultsThe major source of aerosol loss was at the nasal interface (∼25%). Significant differences in deposition on circuit components were seen between nebulizers. The nebulizer residual was higher for BEJN (P = .006), and circuit losses, including the humidifier, were higher for vibrating mesh nebulizer (P = .006). There were no differences in delivery to the filter and head model. For 60 L/min gas flow, as infusion pump flow was increased, the rate of aerosol delivery (CPM/min) increased, for BEJN from 338 to 8,111; for vibrating mesh nebulizer, maximum delivery was 2,828.ConclusionsThe model defined sites of aerosol losses during continuous nebulization and provided a realistic in vitro system for testing aerosol delivery during continuous nebulization. Real-time analysis can quantify effects of multiple changes in variables (nebulizer technology, infusion rate, gas flow, and ventilation) during a given experiment.Copyright © 2023 by Daedalus Enterprises.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.