Conf Proc IEEE Eng Med Biol Soc · Jan 2013
Demonstrating the accuracy of an in-hospital ambulatory patient monitoring solution in measuring respiratory rate.
- N Donnelly, T Hunniford, R Harper, A Flynn, A Kennedy, D Branagh, and J McLaughlin.
- Conf Proc IEEE Eng Med Biol Soc. 2013 Jan 1;2013:6711-5.
AbstractThis paper presents clinical testing conducted to evaluate the accuracy of Aingeal, a wireless in-hospital patient monitor, in measuring respiration rate via impedance pneumography. Healthy volunteers were invited to simultaneously wear a CE Marked Aingeal vital signs monitor and a capnograph, the current gold standard in respiration rate measurement. During the test, participants were asked to undergo a series of defined breathing protocols which included normal breathing, paced breathing between 8-23 breaths per minute (bpm) and a recovery period following moderate exercise. Statistical analysis of the data collected shows a mean difference of -0.73, a standard deviation of 1.61, limits of agreement of -3.88 and +2.42 bpm and a P-value of 0.22. This testing demonstrates comparable performance of the Aingeal device in measuring respiration rate with a well-accepted and widely used alternative method.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.