• J Trauma Acute Care Surg · Jun 2012

    Comparative Study

    Motion generated in the unstable cervical spine during the application and removal of cervical immobilization collars.

    • Mark L Prasarn, Bryan Conrad, Gianluca Del Rossi, MaryBeth Horodyski, and Glenn R Rechtine.
    • Department of Orthopedics, Strong Memorial Hospital, University of Rochester, Rochester, New York 14618, USA. markprasarn@yahoo.com
    • J Trauma Acute Care Surg. 2012 Jun 1;72(6):1609-13.

    BackgroundMany studies have compared the restriction of motion that immobilization collars provide to the injured victim. No previous investigation has assessed the amount of motion that is generated during the fitting and removal process. The purpose of this study was to compare the three-dimensional motion generated when one-piece and two-piece cervical collars are applied and removed from cadavers intact and with unstable cervical spine injuries.MethodsFive fresh, lightly embalmed cadavers were tested three times each with either a one-piece or two-piece cervical collar in the supine position. Testing was performed in the intact state, following creation of a global ligamentous instability at C5-C6. The amount of angular motion resulting from the collar application and removal was measured using a Fastrak, three-dimensional, electromagnetic motion analysis device (Polhemus Inc., Colchester, VT). The measurements recorded in this investigation included maximum values for flexion/extension, axial rotation, medial/lateral flexion, anterior/posterior displacement, axial distraction, and medial/lateral displacement at the level of instability.ResultsThere was statistically more motion observed with application or removal of either collar following the creation of a global instability. During application, there was a statistically significant difference in flexion/extension between the one-piece (1.8 degrees) and two-piece (2.6 degrees) collars, p = 0.009. There was also a statistically significant difference in anterior/posterior translation between the one-piece (3.6 mm) and two-piece (3.4 mm) collars, p = 0.015. The maximum angulation and displacement during the application of either collar was 3.4 degrees and 4.4 mm. Statistical analysis revealed no significant differences between the one-piece and two-piece collars during the removal process. The maximum angulation and displacement during removal of either collar type was 1.6 degrees and 2.9 mm.ConclusionsThere were statistically significant differences in motion between the one-piece and two-piece collars during the application process, but it was only 1.2 degrees in flexion/extension and 0.2 mm in anterior/posterior translation. Overall, the greatest amount of angulation and displacement observed during collar application was 3.4 degrees and 4.4 mm. Although the exact amount of motion that could be deleterious to a cervical spine-injured patient is unknown, collars can be placed and removed with manual in-line stabilization without large displacements. Only trained practitioners should do so and with great care given that some motion in all planes does occur during the process.Copyright © 2012 by Lippincott Williams & Wilkins.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…