Comparative Study Observational StudyComparison of the accuracy of hemoglobin point of care testing using HemoCue and GEM Premier 3000 with automated hematology analyzer in emergency room.
- Jan Zatloukal, Jiri Pouska, Jakub Kletecka, Richard Pradl, and Jan Benes.
- Department of Anesthesiology and Intensive Care, The Faculty of Medicine, The University Hospital, Plzen - Charles University Prague, alej Svobody 80, 304 60, Plzeň, Czech Republic. zatloukalj@fnplzen.cz.
- J Clin Monit Comput. 2016 Dec 1; 30 (6): 949-956.
AbstractThe laboratory analysis provides accurate, but time consuming hemoglobin level estimation especially in the emergency setting. The reliability of time-sparing point of care devices (POCT) remains uncertain. We tested two POCT devices accuracy (HemoCue(®)201(+) and Gem(®)Premier™3000) in routine emergency department workflow. Blood samples taken from patients admitted to the emergency department were analyzed for hemoglobin concentration using a laboratory reference Beckman Coulter LH 750 (HBLAB), the HemoCue (HBHC) and the Gem Premier 3000 (HBGEM). Pairwise comparison for each device and HbLAB was performed using correlation and the Bland-Altman methods. The reliability of transfusion decision was assessed using three-zone error grid. A total of 292 measurements were performed in 99 patients. Mean hemoglobin level were 115 ± 33, 110 ± 28 and 111 ± 30 g/l for HbHC, HbGEM and HbLAB respectively. A significant correlation was observed for both devices: HbHC versus HbLAB (r(2) = 0.93, p < 0.001) and HBGEM versus HBLAB (r(2) = 0.86, p < 0.001). The Bland-Altman method revealed bias of -3.7 g/l (limits of agreement -20.9 to 13.5) for HBHC and HBLAB and 2.5 g/l (-18.6 to 23.5) for HBGEM and HBLAB, which significantly differed between POCT devices (p < 0.001). Using the error grid methodology: 94 or 91 % of values (HbHC and HbGEM) fell in the zone of acceptable difference (A), whereas 0 and 1 % (HbHC and HbGEM) were unacceptable (zone C). The absolute accuracy of tested POCT devices was low though reaching a high level of correlation with laboratory measurement. The results of the Morey´s error grid were unfavorable for both POCT devices.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.