• Vet Anaesth Analg · Sep 2012

    Evaluation of a minimally invasive non-calibrated pulse contour cardiac output monitor (FloTrac/Vigileo) in anaesthetized dogs.

    • Rima N Bektas, Annette P N Kutter, Sonja Hartnack, Rahel S Jud, Manuela Schnyder, José M Matos, and Regula Bettschart-Wolfensberger.
    • Section of Anaesthesiology, Equine Department, Vetsuisse Faculty, University of Zurich, Switzerland. rbektas@vetclinics.uzh.ch
    • Vet Anaesth Analg. 2012 Sep 1;39(5):464-71.

    ObjectiveTo evaluate the accuracy of a new cardiac output monitor (FloTrac/Vigileo), originally designed for humans, in dogs. This pulse contour cardiac output monitoring system cannot be calibrated and measures cardiac output (Q̇t) from a standard arterial catheter.Study DesignProspective experimental trial.AnimalsEight adult Beagle dogs weighing 13.1 (9.8-17.1) kg [median (range)].MethodsAnaesthesia in the dogs was maintained using isoflurane. A pulmonary artery catheter and a metatarsal arterial catheter (22 gauge) were placed. Cardiac output was measured simultaneously 331 times by thermodilution and FloTrac technique. A broad spectrum of Q̇t measurements was achieved through alterations of isoflurane concentration, administration of propofol boluses and dobutamine infusions. Agreement between the methods was quantified with Bland Altman analysis and disagreement was assessed with linear mixed models. Results  Median (10th and 90th percentile) cardiac output as measured with thermodilution was 2.54 (1.47 and 5.15) L minute(-1) and as measured with FloTrac 8.6 (3.9 and 17.3) L minute(-1) . FloTrac measurements were consistently higher with a mean bias of 7 L minute(-1) and limits of agreement of -3.15 to 17.17 L minute(-1) . Difference between the methods was most pronounced in high Q̇t measurements. Linear mixed models showed an estimated difference between the two methods of 8.05 (standard error 1.18) L minute(-1) and a significant interaction between mean arterial pressure and method. Standard deviation (4.45 higher) with the FloTrac method compared to thermodilution was increased.ConclusionCompared to thermodilution measurements, the FloTrac system was influenced to a higher degree by arterial blood pressure, resulting in consistent overestimation of cardiac output.Clinical RelevanceThe FloTrac monitor, whose algorithms were developed based on human data, cannot be used as an alternative for thermodilution in dogs.© 2012 The Authors. Veterinary Anaesthesia and Analgesia. © 2012 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesiologists.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.