    Evaluation of aerosol generator devices at 3 locations in humidified and non-humidified circuits during adult mechanical ventilation.

    • Arzu Ari, Hasan Areabi, and James B Fink.
    • Division of Respiratory Therapy, School of Health Professions, Georgia State University, PO Box 4019, Atlanta GA 30302-4019, USA. aari1@gsu.edu
    • Resp Care. 2010 Jul 1;55(7):837-44.

    BackgroundThe position of the jet or ultrasonic nebulizer in the ventilator circuit impacts drug delivery during mechanical ventilation, but has not been extensively explored, and no study has examined all of the commonly used nebulizers.MethodsDrug delivery from jet, vibrating-mesh, and ultrasonic nebulizers and pressurized metered-dose inhaler (pMDI) with spacer was compared in a model of adult mechanical ventilation, with heated/humidified and non-humidified ventilator circuits. Albuterol sulfate was aerosolized at 3 circuit positions: (1) between the endotracheal tube and the Y-piece; (2) 15 cm from Y-piece; and (3) 15 cm from the ventilator, with each device (n = 3) using adult settings (tidal volume 500 mL, ramp flow pattern, 15 breaths/min, peak inspiratory flow 60 L/min, and PEEP 5 cm H(2)O). The drug deposited on an absolute filter distal to an 8.0-mm inner-diameter endotracheal tube was eluted and analyzed via spectrophotometry (276 nm), and is reported as mean +/- SD percent of total nominal or emitted dose.ResultsThe vibrating-mesh nebulizer, ultrasonic nebulizer, and pMDI with spacer were most efficient in position 2 with both non-humidified (30.2 +/- 1.0%, 24.7 +/- 4.4%, and 27.8 +/- 3.3%, respectively) and heated/humidified circuits (16.8 +/- 2.6%, 16.5 +/- 4.3%, and 17 +/- 1.0%, respectively). In contrast, the jet nebulizer was most efficient in position 3 under both non-humidified (14.7 +/- 1.5%) and heated/humidified (6.0 +/- 0.1%) conditions. In positions 2 and 3, all devices delivered approximately 2-fold more drug under non-humidified than under heated/humidified conditions (P < .01). At position 1 only the pMDI delivered substantially more drug than with the non-humidified circuit.ConclusionDuring mechanical ventilation the optimal drug delivery efficiency depends on the aerosol generator, the ventilator circuit, and the aerosol generator position.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.