• Resp Care · Sep 2004

    Albuterol delivery from a metered-dose inhaler with spacer is reduced following short-duration manual ventilation in a neonatal ventilator-lung model.

    • Ralph A Lugo and Julie Ballard.
    • Department of Pharmacotherapy and Pediatrics, University of Utah College of Pharmacy and School of Medicine, Salt Lake City, 84112-5820. rlugo@pharm.utah.edu.
    • Resp Care. 2004 Sep 1;49(9):1029-34.

    IntroductionAlbuterol aerosol is commonly administered to mechanically ventilated neonates via metered-dose inhaler (MDI) with spacer. The spacer increases the dead space in the ventilation circuit, and some institutions limit the amount of time the spacer remains in line, to minimize carbon dioxide retention and the risk of hypercarbia. However, minimizing the amount of time the spacer remains in line might also limit albuterol delivery to the patient.ObjectiveTo determine whether limiting the amount of time the spacer is left in line after MDI actuation significantly reduces albuterol delivery.MethodsWe conducted a bench study with a neonatal ventilator-lung model that included a Bird VIP ventilator, in a time-cycled, pressure-limited, continuous-flow mode, with settings to simulate a 1-kg infant with moderate lung disease: peak inspiratory pressure 25 cm H2O, positive end-expiratory pressure 4 cm H2O, respiratory rate 30 breaths/min, inspiratory time 0.35 s, tidal volume approximately 7 mL. The circuit was attached to a 3.0-mm inner-diameter endotracheal tube and a neonatal test lung. We tested 5 methods of MDI albuterol administration. The first 3 methods used a spacer attached to the ETT and either 5, 15, or 30 manual breaths (flow 6 L/min, respiratory rate 30 breaths/min, peak inspiratory pressure 25 cm H2O) were delivered after each MDI actuation (2 actuations). The final 2 methods used an in-line spacer (placed between the circuit Y-piece and the endotracheal tube) with the spacer kept in line for 30 or 60 s after each actuation (2 actuations). A breathing filter was placed between the ETT and test lung to trap the aerosolized albuterol.ResultsMean +/- SD albuterol delivery was 2.3 +/- 0.5%, 3.6 +/- 1.8%, and 5.1 +/- 1.3% after 5, 15, and 30 manual breaths, respectively (p < or = 0.05 for 30 breaths vs 5 and 15 breaths). Albuterol delivery was 3.7 +/- 1.3% when the spacer was left in line for 30 s, versus 3.7 +/- 0.6% when it was left in line for 60 s.ConclusionsLimiting the time that the spacer was left in line after each MDI actuation significantly reduced albuterol delivery in our neonatal ventilator-lung model.Copyright 2004 Daedalus Enterprises

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.