Accuracy of tidal volume, compliance, and resistance measurements on neonatal ventilator displays: an in vitro assessment.
- Soraya Abbasi, Emidio Sivieri, Robin Roberts, and Haresh Kirpalani.
- Division of Neonatology, Children's Hospital of Philadelphia, University of Pennsylvania, Philadelphia, PA, USA. abbasis@uphs.upenn.edu
- Pediatr Crit Care Me. 2012 Jul 1;13(4):e262-8.
ObjectiveTo determine the accuracy of measures of respiratory mechanics derived from neonatal ventilators using an in vitro passive physical lung model to simulate newborn pulmonary conditions.DesignTest lung models.SettingLaboratory-based measurements.InterventionsThree test lungs were constructed to simulate three severities of neonatal lung disease, with ranges of compliance from 0.5 to 2.0 mL/cm H2O and resistance from 25 to 150 cm H2O/(L/sec). Each ventilator was tested using 27 combinations of peak inspiratory pressure (15-25 cm H2O), positive end-expiratory pressure (5-7 cm H2O), and rate settings (20-60 B/min). Data were compared for five different ventilators across simulated lung severity as the ratio of ventilator readout to test lung reference value. A ratio of 1.0 indicated a completely unbiased result.Main ResultsOverall, four of the five ventilators under-read expired tidal volume by about 1%-12% across all lung conditions, whereas the VIP Bird readout ranged from -4% to +4% bias. Changes in ventilator settings had only a modest effect on mechanics readout. As peak inspiratory pressure progressed from 15 to 25 cm H2O, bias in tidal volume readout changed from +5.0% to -2.5% (p < .001) in the VIP Bird, and from -11% to -9% (p < .001) in the Draeger Babylog VN500. Between positive end-expiratory pressure levels of 5 and 7 cm H2O, tidal volume bias in the Babylog varied between -13% and -7% (p < .001). In progressing from simulated normal to severely ill lung condition, bias in compliance measurements by the Avea and SLE5000 increased from -18% to -40% whereas in the VIP Bird it remained between -17% to -13%, and in the Draeger Evita XL-neo it changed from +17% to -13% and from -8% to -16% in the Babylog. Ratio of ventilator resistance readout to reference value with progressing simulated lung condition changed from 2.0 to 1.0 for the Draeger Evita, 1.6 to 1.1 for the Babylog, 4.2 to 2.0 for the SLE, and from 11.7 to 5.6 for the VIP Bird. The Avea, by design, did not display resistances >100 cm H2O/(L/sec), but overestimated the simulated normal lung resistance of 25 cm H2O/(L/sec) by a factor of 2.5.ConclusionsNeonatal ventilator respiratory mechanics measurements and computation methods need further standardization to be useful in clinical settings.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.