• J Clin Anesth · May 1994

    Postanesthesia monitoring revisited: frequency of true and false alarms from different monitoring devices.

    • L Wiklund, B Hök, K Ståhl, and A Jordeby-Jönsson.
    • Department of Anesthesiology and Intensive Care, Uppsala University Hospital, Sweden.
    • J Clin Anesth. 1994 May 1;6(3):182-8.

    Study ObjectivesTo determine the frequency of true and false alarms and to determine the frequency of alarm failures for various parameters when using a postanesthesia monitoring system.DesignOpen prospective study.SettingPostanesthesia care unit at a university hospital.Patients123 ASA physical status I-IV patients who underwent general or spinal-epidural anesthesia for general, urologic, orthopedic, ophthalmologic, otolaryngologic, or plastic surgery.Measurements And Main ResultsMonitoring included ECG, oxygen saturation, and respiratory rate (RR) by either transthoracic impedance or nasal-oral air-flow detection. We undertook careful, continuous observation of patients and monitors, recording of true and false alarms; the latter were defined as those coinciding with a clearly observed, unrelated cause. During a mean observation period of 101 minutes per patient, the average frequency of pulse oximetry alarms was once every 8 minutes, with 77% of the alarms being false, caused by sensor displacement, motion artifacts, poor perfusion, or a combination of these factors. Apnea alarms occurred on average once every 37 minutes, with the "false" fraction being 28% and 27% for impedance and flow detection, respectively. The impedance sensor failed to detect apnea on at least 6 occasions; the flow sensor failed on 1 occasion. The coincidence of pulse oximetry and apnea alarms was small, and ECG exhibited a low alarm rate but a high fraction of false alarms. Patients receiving opioids and neuromuscular relaxants had a higher frequency of "true" apneas than patients not receiving these drugs. No clear correlation was found in regard to age distribution, type of surgery, duration of anesthesia, oxygen administration, or ASA physical status between the patient groups that exhibited many apneas or many pulse oximeter alarms and the whole study population.ConclusionsThe high frequency of apnea strongly motivates the use of continuous RR monitoring, preferably by flow-sensing techniques, since both central and obstructive apneas are then detected. Further study and development is necessary before pulse oximetry can be unconditionally recommended for postanesthesia monitoring.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.