• J Perinatol · Sep 2002

    Comparative Study Clinical Trial Controlled Clinical Trial

    Continuous blood gas monitoring using an in-dwelling optode method: comparison to intermittent arterial blood gas sampling in ECMO patients.

    • K Rais-Bahrami, Oswaldo Rivera, Gerald T Mikesell, and Billie L Short.
    • Department of Neonatology, Children's National Medical Center and The George Washington University School of Medicine, Washington, DC 20010, USA.
    • J Perinatol. 2002 Sep 1;22(6):472-4.

    IntroductionThe ability to measure postmembrane arterial blood gases is essential in the management of critically ill neonates treated with extracorporeal membrane oxygenation (ECMO). A new technology using, the Paratrend 7 system (Diametrics Medical, High Wycombe,UK) allows for continuous measurement of pH, PCO(2) and PO(2), and calculates oxygen saturation, bicarbonate, and base excess.ObjectiveTo evaluate and compare the results of continuous blood gas measurement using the Paratrend 7 system with a standard system of blood gas analysis in our intensive care unit.DesignProspective, controlled, interventional study.SettingThe neonatal intensive care unit of a tertiary referral center.PatientsNeonates who required extracorporeal life support and were expected to have frequent postmembrane arterial blood sampling during the testing period.Research Design And MethodTo enable Paratrend 7 sensor access to the ECMO circuit, the postmembrane access port extension set that is routinely used for blood drawn for blood gas analysis was used. The study began with the insertion of the Paratrend 7 sensor. Subjects remained on the study until the ECMO was discontinued and/or frequent blood gases were no longer needed. The blood gas results from the Paratrend 7 system were not used in clinical management of the patient. BLOOD GAS MEASUREMENT: During the study period, with each blood sample drawn for laboratory analysis, a printout from the Paratrend 7 monitor was recorded for comparison.ResultsA total of 242 pairs of blood gas samples were collected from 10 neonates. The mean bias/precision for pH was -0.02/0.04; for PO(2) 68.35/93.44 mm Hg; and for PCO(2) 1.75/4.23 mm Hg. The correlation (r value) between the sensor reading and the blood gases were 0.89 for pH, 0.96 for PO(2), and 0.73 for PCO(2) (Table 1).ConclusionThe blood gases compared in the two methods had a strong correlation for pH, PCO(2) and PO(2). Results of this study indicate that this technology provides an accurate means of monitoring continuous blood gas parameters in neonatal ECMO patients. Use of the Paratrend 7 should allow reduced health-care provider exposure to blood and decreased patient iatrogenic blood loss.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…