    Transcutaneous monitoring of partial pressure of carbon dioxide in the elderly patient: a prospective, clinical comparison with end-tidal monitoring.

    • Andrea Casati, Grazia Squicciarini, Giulia Malagutti, Marco Baciarello, Marta Putzu, and Andrea Fanelli.
    • Department of Anesthesia and Pain Therapy, University of Parma, and Ospedale Maggiore di Parma, via Gramsci 14, 43100 Parma, Italy. acasati@ao.pr.it
    • J Clin Anesth. 2006 Sep 1;18(6):436-40.

    Study ObjectiveTo evaluate the accuracy and precision of estimation of partial pressure of carbon dioxide (Pa(CO2)) using end-tidal or transcutaneous CO2 (TcP(CO2)) measurements during mechanical ventilation in the elderly patient.DesignA prospective, observational study was conducted.SettingsThe study was done in the anesthesia department of a university hospital.PatientsSeventeen anesthetized, mechanically ventilated patients older than 60 years were studied.Interventions And MeasurementsDuring standard sevoflurane anesthesia, and after proper calibration and an equilibration time of 30 minutes with stable hemodynamic and respiratory variables, arterial (Pa(CO2)), end-tidal (Pet(CO2)), and transcutaneous (TcP(CO2)) CO2 partial pressures were determined. In each patient, 1 to 5 sample sets (Pa(CO2), Pet(CO2), and TcP(CO2)) were obtained.Main ResultsA total of 45 sample sets were obtained from the patients studied. The Pa(CO2) values ranged between 21 and 58 mm Hg. The Pa(CO2) - Pet(CO2) tension gradient was 6 +/- 5 mmHg (95% confidence interval, -3 to 16 mmHg), whereas the Pa(CO2) - TcP(CO2) tension gradient was 2 +/- 4 mmHg (95% confidence interval, -6 to 9 mmHg) (P = 0.0005). The absolute value of the difference between Pa(CO2) and Pet(CO2) was 3 mm Hg or less in 7 of 45 sample sets (15%), whereas the absolute value of the difference between Pa(CO2) and TcP(CO2) was 3 mm Hg or less in 21 of 45 sample sets (46%) (P = 0.003). Linear regression analysis for TcP(CO2) versus Pa(CO2) showed a slope of 0.84 (r(2) = 0.73), whereas the linear regression analysis for Pet(CO2) versus Pa(CO2) showed a slope of 0.54 (r(2) = 0.50).ConclusionTranscutaneous monitoring of CO(2) partial pressure gives a more accurate estimation of arterial CO(2) partial pressure than does Pet(CO2) monitoring.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.